The Mass Spectrometer

A concise revision guide to how a mass spectrometer works for AQA A Level Chemistry, including vaporisation, electron impact ionisation, acceleration, deflection, detection and the interpretation of m/z values.

Paper 1
AQA
3.1.1 Atomic Structure
7405/1
AQA specification3 spec points in this lesson
  • 3.1.1.2iv-a
  • 3.1.1.2iv-b
  • 3.1.1.2iv-d
What these spec points say
  • 3.1.1.2iv-a understand ionisation in a simple time of flight (TOF) mass spectrometer
  • 3.1.1.2iv-b understand acceleration in a simple time of flight (TOF) mass spectrometer, including giving all ions constant kinetic energy
  • 3.1.1.2iv-d understand ion detection in a simple time of flight (TOF) mass spectrometer

Mass Spectrometer: Follow the Ions

Choose a step to follow the ions through the instrument: ionisation, acceleration, deflection, detection and changing the magnetic field.

Drag to rotate
Step
View

Hydrogen atom or H⁺ ion Electron Ion with smaller m/z Ion with larger m/z N pole S pole

© Dr. Mohammed Al-Fatah – onlinelearningsystem.net

1

Before You Start

This page builds on GCSE atomic structure: isotopes, ions and the forces between charges.

  • Isotopes are atoms of the same element with the same number of protons but different numbers of neutrons.
  • An atom that loses an electron becomes a positive ion.
  • Opposite charges attract and like charges repel.
  • The relative atomic mass of an element is a mean that allows for the abundance of each isotope.
Before you start

GCSE Recap: Isotopes, Ions and Charges

Before you read on, check the GCSE ideas about isotopes, ions and charges that the instrument relies on.

2

What a Mass Spectrometer Does

AQA note: AQA examines only the time of flight mass spectrometer. The magnetic deflection instrument on this page is background reading and is not examined by AQA.

Ionisation, m/z and the detector current still apply, but learn the stages from the Time of Flight page.

A mass spectrometer separates positive ions according to their mass-to-charge ratio, written as m/z.

It is used to identify isotopes, compare their relative abundance, and provide data used to calculate relative atomic mass. The sample must be converted into particles that can move through the machine and be detected as ions.

The main sequence is: vaporisation, ionisation, acceleration, deflection and detection.

Key idea: The mass spectrometer does not directly weigh atoms. It forms positive ions, separates them by m/z, and records detector signals that show relative abundance.

Labelled electron bombardment mass spectrometer with the five stages vaporisation, ionisation, acceleration, deflection and detection, and both ion paths reaching the detector

The detector records ions that pass through the selected path, allowing the computer to build a spectrum from their m/z values and abundances.

3

Why the Instrument Uses a Vacuum

The mass spectrometer operates at low pressure and under a vacuum.

This matters because air particles would otherwise collide with the ions, change their path, or become ionised themselves. If air particles reached the detector, they could produce extra signals and reduce the reliability of the spectrum.

Exam focus: A vacuum prevents collisions with air particles and prevents air particles from being ionised and detected.

Check your understanding

Quick Check: What If Air Leaks In?

Explain what would go wrong if air leaked into the instrument.

4

Stage 1: Vaporisation and Ionisation

The sample must be in the gaseous state so that its particles can move through the mass spectrometer. If the sample is not already gaseous, it is first vaporised.

In electron impact ionisation, an electron gun fires high-energy electrons at the gaseous atoms or molecules. These electrons collide with the sample and remove an outer electron, forming a positive ion.

For an atom X, this can be represented as: X(g) + e– → X+(g) + 2e–.

Remember: Ionisation produces positive ions because an electron has been removed.

Check your understanding

Quick Check: Write the Ionisation Equation

Choose the correct ionisation equation for a molecule that is not on this page.

5

Stage 2: Acceleration

The positive ions are accelerated by an electric field. The field pulls the positive ions towards a negatively charged plate or region.

This forms a narrow beam of ions travelling through the instrument. Ions with the same charge are given the same kinetic energy as they are accelerated into the flight path.

Exam focus: Positive ions accelerate towards the negative plate because opposite charges attract.

Check your understanding

Quick Check: What Gets Accelerated?

Decide whether the statement about un-ionised atoms is true or false.

6

Stage 3: Deflection

The beam of positive ions passes through a magnetic field. The magnetic field deflects the ions so that their path bends.

The amount of deflection depends on the mass-to-charge ratio, m/z. Ions with a smaller m/z are deflected more. Ions with a larger m/z are deflected less.

If the ions all have a charge of 1+, then lighter ions have a smaller m/z and are deflected more than heavier ions. If the charge is different, m/z must be considered directly rather than mass alone.

A 2+ ion forms when a second electron is knocked out during electron impact ionisation.

Its m/z is half that of the 1+ ion of the same mass, so it is deflected more. Most ions are 1+ because removing a second electron needs much more energy.

Ion propertyEffect on m/zEffect on deflection
Lower mass with the same chargeSmaller m/zMore deflection
Higher mass with the same chargeLarger m/zLess deflection
Higher positive charge with the same massSmaller m/zMore deflection

Example: An ion with mass 28 and charge 1+ has m/z = 28. An ion with mass 56 and charge 2+ also has m/z = 28, so both can be detected under the same magnetic field setting.

Smaller m/z means more deflection: a 2+ ion has half the m/z of the 1+ ion of the same mass.

Check your understanding

Quick Check: Compare the m/z Values

Compare the m/z values and deflections of ions that are not on this page.

7

Stage 4: Detection

The deflected ions reach a detector plate. When a positive ion hits the detector, it gains an electron and becomes neutral.

Electrons flow from the detector to the positive ions, creating a small electrical current. This signal is amplified and sent to a computer.

The size of the detector signal is proportional to the number of ions reaching the detector. This is why the peak height or intensity in a mass spectrum shows relative abundance.

Exam focus: More ions hitting the detector produce a larger current, so the signal represents greater abundance.

Check your understanding

Quick Check: Reading the Detector Current

Use the detector currents to compare two isotopes.

8

Changing the Magnetic Field

For a fixed magnetic field strength, only ions with a particular m/z value follow the correct path through the slit and reach the detector.

The magnetic field strength is then changed so that ions with different m/z values can be detected.

Ions with a larger m/z are deflected less, so a stronger magnetic field is needed to bend them onto the detector. A weaker field brings ions with a smaller m/z onto the detector.

The computer combines these detector signals to produce the mass spectrum.

Key idea: Changing the magnetic field lets the instrument scan across different m/z values.

Check your understanding

Quick Check: Scanning to a Heavier Ion

Decide how the magnetic field must change to detect a heavier ion.

Check your understanding

Quick Check: Mass Spectrometer Round-Up

Pick the one fully correct statement in each round to pull the whole page together.

9

Common Exam Points

Mass spectrometry questions often test precise sequencing, the role of the vacuum, the meaning of m/z, and the link between detector current and abundance.

Use the correct sequence

Vaporise the sample if needed, ionise it, accelerate the positive ions, deflect them in a magnetic field, then detect them.

State why a vacuum is needed

A vacuum prevents collisions with air particles and prevents air particles from becoming ionised and detected.

Do not use mass alone when charge can vary

Deflection depends on m/z, not only mass. A higher charge lowers the m/z value.

Link current to abundance

A larger current is produced when more ions reach the detector, so signal intensity is proportional to abundance.

FAQs

These questions secure the wording examiners expect on ionisation, m/z and detection.

What does a mass spectrometer actually measure?

It does not weigh atoms. It separates positive ions by their mass-to-charge ratio, m/z, and records how many ions of each m/z reach the detector.

How does electron impact ionisation form positive ions?

An electron gun fires high-energy electrons at the gaseous sample. Each impact knocks out an electron: X(g) + e− → X+(g) + 2e−.

Why must the instrument be kept under a vacuum?

Air particles would collide with the ions and knock them off course. They could also be ionised themselves and give extra peaks at the detector.

Why are most ions 1+ rather than 2+?

Removing a second electron needs much more energy, so few 2+ ions form. A 2+ ion has half the m/z of the 1+ ion of the same mass.

Do I need to learn this magnetic instrument for the exam?

No. The magnetic deflection stage is background reading. Learn the stages from the Time of Flight page; ionisation, m/z and the detector current apply to both instruments.