The Mass Spectrometer
A concise revision guide to how a mass spectrometer works for AQA A Level Chemistry, including vaporisation, electron impact ionisation, acceleration, deflection, detection and the interpretation of m/z values.
What a Mass Spectrometer Does
AQA note: AQA examines only the time of flight mass spectrometer. The magnetic deflection instrument on this page is background reading and is not examined by AQA. Ionisation, m/z and the detector current still apply, but learn the stages from the Time of Flight page.
A mass spectrometer separates positive ions according to their mass-to-charge ratio, written as m/z.
It is used to identify isotopes, compare their relative abundance, and provide data used to calculate relative atomic mass. The sample must be converted into particles that can move through the machine and be detected as ions.
The main sequence is: vaporisation, ionisation, acceleration, deflection and detection.
Key idea: The mass spectrometer does not directly weigh atoms. It forms positive ions, separates them by m/z, and records detector signals that show relative abundance.
The detector records ions that pass through the selected path, allowing the computer to build a spectrum from their m/z values and abundances.
Why the Instrument Uses a Vacuum
The mass spectrometer operates at low pressure and under a vacuum.
This matters because air particles would otherwise collide with the ions, change their path, or become ionised themselves. If air particles reached the detector, they could produce extra signals and reduce the reliability of the spectrum.
Exam focus: A vacuum prevents collisions with air particles and prevents air particles from being ionised and detected.
Quick Check: What If Air Leaks In?
Write a short explanation, then compare it with the mark points and the model answer.
Stage 1: Vaporisation and Ionisation
The sample must be in the gaseous state so that its particles can move through the mass spectrometer. If the sample is not already gaseous, it is first vaporised.
In electron impact ionisation, an electron gun fires high-energy electrons at the gaseous atoms or molecules. These electrons collide with the sample and remove an outer electron, forming a positive ion.
For an atom X, this can be represented as: X(g) + e– → X+(g) + 2e–.
Remember: Ionisation produces positive ions because an electron has been removed.
Quick Check: Write the Ionisation Equation
Choose the equation that shows a water molecule being ionised by electron bombardment.
Stage 2: Acceleration
The positive ions are accelerated by an electric field. The field pulls the positive ions towards a negatively charged plate or region.
This forms a narrow beam of ions travelling through the instrument. Ions with the same charge are given the same kinetic energy as they are accelerated into the flight path.
Exam focus: Positive ions accelerate towards the negative plate because opposite charges attract.
Quick Check: What Gets Accelerated?
Decide whether the statement about atoms that were not ionised is true or false.
Stage 3: Deflection
The beam of positive ions passes through a magnetic field. The magnetic field deflects the ions so that their path bends.
The amount of deflection depends on the mass-to-charge ratio, m/z. Ions with a smaller m/z are deflected more. Ions with a larger m/z are deflected less.
If the ions all have a charge of 1+, then lighter ions have a smaller m/z and are deflected more than heavier ions. If the charge is different, m/z must be considered directly rather than mass alone.
A 2+ ion forms when a second electron is knocked out during electron impact ionisation. Its m/z is half that of the 1+ ion of the same mass, so it is deflected more. Most ions are 1+ because removing a second electron needs much more energy.
| Ion property | Effect on m/z | Effect on deflection |
|---|---|---|
| Lower mass with the same charge | Smaller m/z | More deflection |
| Higher mass with the same charge | Larger m/z | Less deflection |
| Higher positive charge with the same mass | Smaller m/z | More deflection |
Example: An ion with mass 28 and charge 1+ has m/z = 28. An ion with mass 56 and charge 2+ also has m/z = 28, so both can be detected under the same magnetic field setting.
Quick Check: Compare the m/z Values
Work out m/z for each ion, then answer five quick questions on deflection.
Stage 4: Detection
The deflected ions reach a detector plate. When a positive ion hits the detector, it gains an electron and becomes neutral.
Electrons flow from the detector to the positive ions, creating a small electrical current. This signal is amplified and sent to a computer.
The size of the detector signal is proportional to the number of ions reaching the detector. This is why the peak height or intensity in a mass spectrum shows relative abundance.
Exam focus: More ions hitting the detector produce a larger current, so the signal represents greater abundance.
Quick Check: Reading the Detector Current
Use two detector readings to make a deduction about a boron sample.
Changing the Magnetic Field
For a fixed magnetic field strength, only ions with a particular m/z value follow the correct path through the slit and reach the detector.
The magnetic field strength is then changed so that ions with different m/z values can be detected. Ions with a larger m/z are deflected less, so a stronger magnetic field is needed to bend them onto the detector. A weaker field brings ions with a smaller m/z onto the detector. The computer combines these detector signals to produce the mass spectrum.
Key idea: Changing the magnetic field lets the instrument scan across different m/z values.
Quick Check: Scanning to a Heavier Ion
Decide how the magnetic field must change to bring a heavier ion onto the detector.
Quick Check: Follow a Sample Through the Instrument
Put the seven events in the order in which they happen.
Common Exam Points
Mass spectrometry questions often test precise sequencing, the role of the vacuum, the meaning of m/z, and the link between detector current and abundance.
Use the correct sequence
Vaporise the sample if needed, ionise it, accelerate the positive ions, deflect them in a magnetic field, then detect them.
State why a vacuum is needed
A vacuum prevents collisions with air particles and prevents air particles from becoming ionised and detected.
Do not use mass alone when charge can vary
Deflection depends on m/z, not only mass. A higher charge lowers the m/z value.
Link current to abundance
A larger current is produced when more ions reach the detector, so signal intensity is proportional to abundance.
QuickSnap
In a mass spectrometer, the sample is first vaporised if needed, then ionised by electron impact ionisation to form positive ions. The positive ions are accelerated by an electric field and deflected by a magnetic field according to their m/z value. Ions with smaller m/z values are deflected more, while ions with larger m/z values are deflected less. When ions reach the detector, they gain electrons and produce a current. The size of this current is proportional to the abundance of the ions reaching the detector.
Ionisation
High-energy electrons remove electrons from gaseous particles, forming positive ions.
Acceleration
Positive ions are attracted towards a negative electric field and form a beam.
Deflection
The magnetic field separates ions by m/z. Smaller m/z means greater deflection.
Detection
Positive ions gain electrons at the detector, producing a current proportional to abundance.
Master Atomic Structure and the Periodic Table for AQA A Level Chemistry
Continue from these free revision notes into the full 3.1.1 and 3.2.1 Atomic Structure and the Periodic Table course. The five guided video lessons are ready now; the AQA MCQ bank, teacher-marked short-answer questions, supplementary lessons and KASP spec-point mapping are being completed.
Guided video teaching
Learn the chemistry and exam technique through structured video lessons with worked examples and walkthroughs.
Instant MCQ feedback
Auto-marked MCQ quizzes provide immediate diagnostic feedback for every answer choice.
Teacher-marked SAQs
Submit written exam responses and receive chemistry specialist feedback with improvement guidance.
Progress tracking
Identify strengths and weaknesses across the full 3.1.1 and 3.2.1 specification with targeted reporting.
See how the course works
Click play to start the course preview animation.
Some ionic radii are shown.
| Ion | Ionic radius / nm |
|---|---|
| Na+ | 0.102 |
| K+ | 0.138 |
| F− | 0.133 |
| Cl− | 0.180 |
Which compound has the strongest ionic bonding?
Explain why the metallic bonding in magnesium is much stronger than that in sodium.
Copyright notice: This OLS revision content, including the explanations, layout, diagrams, tables and embedded learning structure, is authored for Online Learning System by Dr. Mohammed Al-Fatah. It may not be copied, reproduced, redistributed or adapted without written permission.
Keep this note — free
Save your progress across every AQA topic. A free account remembers which topics you have covered, saves your question scores, and syncs across your phone and laptop.
- Track every topic you have finished
- Keep your practice-question scores
- No payment, no card, free forever
Already registered? Log in